NMPA Issues Classification Guidelines for Ultrasound Products Intended for Medical Aesthetic Use
On November 23, 2026, the National Medical Products Administration (NMPA) issued the Guiding Principles for the Classification of Ultrasound Products for Medical Aesthetic Use (Announcement No. 43 of 2026), providing specific requirements for determining the regulatory status and classification of ultrasound products intended for medical aesthetic applications. Effective from the date of issuance, the Guiding Principles aim to strengthen the supervision of these products and promote the orderly development of the medical aesthetics industry. They clarify the circumstances under which ultrasound products are regulated as medical devices and establish classification criteria based on their operating principles, technical parameters, and intended uses.
Determination of Medical Device Regulatory Status
The Guiding Principles specify that the regulatory status of ultrasound products for medical aesthetic use should be determined by comprehensively considering their operating principles and intended uses. Products that meet the definition of a medical device under the Regulations on the Supervision and Administration of Medical Devices and use focused or non-focused ultrasound, through thermal and/or mechanical effects, to achieve medical aesthetic purposes must be regulated as medical devices.
These intended purposes include fat emulsification or dissolution, skin scar repair, improvement of skin laxity and sagging, cheek enhancement, facial contour lifting, skin and contour tightening, body or facial shaping, pore reduction, improvement of skin elasticity, wrinkle reduction, and reduction of skin pigmentation. The inclusion of these claims within a product's intended use is an important factor in determining whether the product falls under medical device regulation.
In contrast, ultrasound products that do not meet the regulatory definition of a medical device are not subject to medical device regulation. However, such products must not claim medical device-related intended uses. Examples include products used solely to enhance the non-invasive penetration of cosmetic essences (excluding penetration enhancement for drugs or medical devices), remove substances from pores for skin cleansing, provide skin warming or physical massage, promote skin relaxation, or remove dead skin cells.
Classification of Focused Ultrasound Products
For focused ultrasound products regulated as medical devices, the Guiding Principles establish three classification scenarios based on intended use and technical parameters.
Ultrasound aesthetic devices intended to ablate fat and achieve fat emulsification or dissolution are classified as Class III medical devices under classification code 01-01.
Other focused ultrasound aesthetic devices are also classified as Class III medical devices under classification code 09-06 if their intended uses include one or more of the following: skin scar repair, improvement of skin laxity or sagging, cheek enhancement, facial contour lifting, contour or skin tightening, lifting and shaping, pore reduction, improvement of skin elasticity, or wrinkle reduction. In addition, products are classified as Class III if they meet any one of the following technical criteria: a spatial-peak temporal-average intensity (Ispta) greater than 100 W/cm², a peak rarefactional acoustic pressure (pr) greater than 3 MPa, or a focal plane distance (d) of 1.5 mm or greater.
Focused ultrasound devices intended to reduce skin pigmentation by promoting local blood circulation and accelerating skin metabolism may be classified as Class II medical devices under classification code 09-06, provided that all three technical conditions are met: Ispta is no greater than 100 W/cm², pr is no greater than 3 MPa, and the focal plane distance is less than 1.5 mm.
Classification of Non-Focused Ultrasound Products
For non-focused ultrasound products, the classification is likewise determined by intended use and technical parameters, with two regulatory scenarios.
Non-focused ultrasound aesthetic devices are classified as Class III medical devices under classification code 09-06 if their intended uses include one or more of the specified aesthetic purposes, such as skin scar repair, improvement of skin laxity, facial contour lifting, skin tightening, lifting and shaping, pore reduction, improvement of skin elasticity, or wrinkle reduction. Products are also classified as Class III if their Ispta exceeds 100 W/cm² or their pr exceeds 3 MPa.
By contrast, non-focused ultrasound devices intended to reduce skin pigmentation by promoting local blood circulation and accelerating skin metabolism are classified as Class II medical devices under classification code 09-06 when both technical parameters remain within the specified limits: Ispta is no greater than 100 W/cm² and pr is no greater than 3 MPa.
Notably, the classification criteria differ between focused and non-focused ultrasound products. The focal plane distance threshold is specified for focused ultrasound products but is not included in the classification criteria for non-focused ultrasound products. This distinction highlights the importance of accurately identifying a product's ultrasound operating principle and evaluating the corresponding technical parameters when determining its regulatory classification.
Implications for Medical Device Manufacturers and Registrants
The issuance of the Guiding Principles provides a clearer regulatory framework for ultrasound products used in medical aesthetics. In particular, it establishes more explicit boundaries between products regulated as medical devices and those outside the scope of medical device regulation. It also clarifies the circumstances under which products should be classified as Class II or Class III medical devices.
Manufacturers and applicants should carefully review their products' intended uses, operating principles, and technical specifications against the criteria in the Guiding Principles. Particular attention should be paid to product claims, including whether the product is intended for fat reduction, skin tightening, facial lifting, wrinkle reduction, or other specified medical aesthetic purposes. Technical parameters such as Ispta, peak rarefactional acoustic pressure, and focal plane distance, where applicable, should also be assessed against the relevant classification thresholds.
For products whose regulatory attributes or classifications are affected by the new requirements, medical device registrants must undertake subsequent registration activities in accordance with the Announcement on Matters Related to Adjustments to Medical Device Classification (NMPA Announcement No. 52 of 2026). From December 1, 2029, ultrasound products for medical aesthetic use that have not obtained the required medical device registration certificates under the applicable classifications may no longer be manufactured, distributed, or imported.
The new requirements underscore the importance of aligning product design, technical specifications, intended-use claims, and registration strategies with China's medical device regulatory framework. Companies currently developing or marketing ultrasound-based aesthetic products should assess the potential impact of the Guiding Principles in advance and plan any necessary registration or compliance adjustments within the transition period.